If you use the Tandem Mobi system for therapy management, it is recommended to delete the Tandem t:slim mobile app from your smartphone.
For users managing therapy with the t:slim X2 insulin pump, please note that the Tandem Mobi mobile app cannot be installed on your smartphone if you want full functionality from the Tandem t:slim mobile app. If it is installed:
- The Tandem t:slim mobile app will be disabled
 - You will not be able to view your pump data from the Tandem t:slim mobile app
 - You will not be able to bolus from the Tandem t:slim mobile app
 - No alerts, alarms, reminders, or malfunctions from the t:slim X2 insulin pump will be viewable from the Tandem t:slim mobile app
 - You will be logged out of your Tandem Diabetes Care account
 - Tandem Source will not receive wireless data uploads from the pump
 

If you receive this notification that the mobile app has been disabled:
- Use your t:slim X2 insulin pump for all therapy decisions
 - Deactivation of the mobile app does not impact pump functionality
 - Ensure the pump's sound volume is set to avoid missing alerts and alarms
 - Connect the pump to your computer to upload data to Tandem Source
 
The Tandem t:slim mobile app can be enabled again for use with the t:slim X2 insulin pump after the Tandem Mobi mobile app is removed from your smartphone.
Always refer to the user guide for important safety information and detailed step-by-step instructions on how to troubleshoot t:slim X2 insulin pump alerts and alarms. A full list of alerts and alarms can be accessed in the user guide.